2017-7-20
Smooth appearance design, ingenious color assortment, white backlit LCD screen, oval buttons, easy parameter setting and microprocessor make the operation more convenient.
Main Features
1.70*40mm white backlit LCD screen can show complete parameters like T,A,C,K.
2.Calibrate 0%A and 100%T automatically.
3.Large sample compartment can hold various cells from 5mm to 100mm and meet different test requirements.
4.RS232 output port, optional printer and professional analysis software UV-PRO1.0 with the functions of quantitative analysis.
5. Manual wavelength setting.
6.Save and read out the test data and values of K and B.
Specifications
Model
E-1000V
Display
70*40mm white backlit
LCD
Wavelength
Range
320 - 1020nm
Slit
Width
4nm
Wavelength
Accuracy
±2nm
Wavelength
Repeatability
≤1nm
Photometric
range
0-200%T, -0.3-3.0A, 0-9999C(9999F)
Photometric
Accuracy
0.5%T
Photometric
Repeatability
0.2%T
Stray
light
≤0.15%T@ 360nm
Stability
0.002A@ 500nm
Output
Port
RS232
Light
Source
Tungsten Halogen Lamp
Power
Requirements
AC 110-220V, 50-60HZ
Shipping
Dimensions & Weight
520*450*320mm, 9Kg