Portable Type
Main Features
●Large LCD screen with controllable white backlight.
●Humanized designs like rack and skid proof shell make it more convenient for users.
●Double-parameter and equipped with pH electrode and conductivity electrode, can recognize electrode automatically.
●Built-in microprocessor chip with functions of auto calibration, auto temperature compensation, data storage, function setup, self-diagnosis, auto power off, low voltage display.
●Auto recognition of standard pH buffers and there are three series of standard buffers for option: USA, NIST and China.
●Able to set pH measurement modes of high purity water and pure water with ammonia.
●Auto recognition of 8 standard conductivity buffers and there are two series of standard buffers for option: USA series, China series.
●Adopt advanced conductivity measurement technology, one point calibration of conductivity electrode can meet the measuring accuracy requirements for test 0.5μ S/cm ~ 200mS/cm.
●Able to switch between conductivity, TDS, salinity and resistivity multinomial calculation for TDS and salinity to ensure the conversion precision of the full scale.
●Conductivity mode has the function of non-linear temperature compensation for high purity water.
●IP57 and equipped with calibration solution. There is optional carrying case.
Specifications
PH |
Measuring range |
(-2.00 ~ 19.99) pH |
Resolution |
0.1/0.01 pH |
|
Accuracy |
±0.02 pH |
|
Input current |
≤2×10-12 A |
|
Input impedance |
≥1×1012 Ω |
|
Stability |
±0.01 pH/3h |
|
Temp compensation |
(0 ~ 100)℃ (automatic) |
|
mV |
Measuring range |
-1999mV ~ 0 ~ 1999mV |
Resolution |
1 mV |
|
Accuracy |
±0.1% FS |
|
Conductivity |
Measuring range |
Conductivity:(0.00 ~ 19.99) μS/cm (20.0 ~ 199.9)μS/cm (200~1999)μS/cm (2.00 ~ 19.99)mS/cm (20.0 ~ 199.9)mS/cm TDS: (0 ~ 100) g/L Salinity: (0 ~ 100) ppt Resistivity: (0 ~ 100) MΩ·cm |
Resolution |
0.01/0.1/1μS/cm, 0.01/0.1 mS/cm |
|
Accuracy |
Electrode: ±1.0% FS, Instrument:±1.5%FS |
|
Temp compensation |
(0 ~ 100)℃ (automatic) |
|
Electrode constants |
0.1/1/10 cm-1 |
|
Reference temperature |
25℃, 20℃, 18℃ |
|
Others |
Data storage |
200 groups |
Storage contents |
Series number, measuring value measuring unit and temperature |
|
Power |
Two AA batteries(1.5V×2) |
|
Dimensions & Weight |
160×90×30mm/360g (instrument) 480×300×60mm/1KG (with packing) |
|
Working Condition |
Ambient temperature |
(5 ~ 35)℃ |
Ambient humidity |
≤85% |
|
IP grade |
IP57 |
Standard Configuration
1. P-520 pH/Conductivity meter 1 unit
2. pH electrode 1 pc
3. Conductivity electrode 1 pc
4. Temperature electrode 1 pc
5. pH buffer (pH4.00, pH6.86, pH9.18/50ml) 1 set
6. Backup AA battery 2 pcs
7. Operation manual 1 copy