Today is 2024-12-03 Tuesday,Welcome to this site 

UV-Vis Spectrophotometer

C-7100/7200 Series

Word:[Big][Middle][Small] QR Code 2017/7/20     Viewed:    

Steady, modern and elegant appearance design. Adopt the newest microcomputer technology and electronic control system. Optimized optical system and structure can both extend new functions and ensure the accuracy, stability and durability.


Main Features

1. 7 inch color LCD screen and long life, more comfortable and sensitive silicone buttons. The instrument can show various scanning curves and charts for users to complete various tests without computers.

2. Support USB storage and different data formats such as Excel, txt and image(PC software). Users can output test data to flash memory, open and edit them on computers directly without any auxiliary software.

3. Advanced hardware and 32-bit Cortex_M3 processor with the clock speed 120MHz. The equipment can store 5000 pieces of data and 500 curves.

4. High-efficiency holographic grating of 1200 lines/mm and low stray light.

5. The equipment has long-life socket type tungsten-halogen and deuterium lamps which can work up to 2000 hours, can switch the lamps according to test needs and record its working time automatically. Socket type lamps make the replacement much easier.

6. Excellent silicon photodiode can guarantee the equipment is highly sensitive and stable.

7. Huge sample chamber and various accessories can meet all kinds of needs.

8. Can be connected to printer directly and output test charts and data.

9. Powerful PC software.

10. Standard RS232, USB(A), USB(B) port for printer, data transfer and PC connection.


UI Design

1. Photometry

There are three test modes: Absorbance, transmittance and energy. 


2. Quantitative test.

To test sample solution concentration, you can choose different methods like coefficient, standard curve, linearity, linear zero crossing and quadratic. Up to 15 standard samples can be used to build a curve. Advance arithmetic make curvilinear regression more precise and test data more accurate.


3. Kinetics Measurement (Time Scanning)

To test the sample chemical reaction process by fixed time scanning the sample solution with fixed wavelength. The equipment can calculate its changing rate after entering the corresponding parameters.


4. Wavelength Scanning (Qualitative Test)

To test sample solution absorbance peak, can scan the sample characteristic curve of any wavelength range between 190 and 1100nm. You can look up the peak value on the device standalone. 


5. Multi wavelength testing

It is much more convenient for users to test the absorbance of several wavelengths for the same sample solution, which is much simpler than single wavelength testing.


6. DNA/Protein Measurement

It is a special function for specific users to make the operation easier. There are two test modes.


Specifications

Model

C-7100

C-7100S

C-7100A

C-7200

C-7200S

C-7200A

Display

7 inch LCD screen

Keyboard Control

Silicone Button

Optical System

Single Beam

Double Beam

Holographic grating, 1200 lines/mm

Spectral Bandwidth

2nm

1nm

0.5,1,2,4nm

2nm

1nm

0.5,1,2,4nm

Wavelength Range

190 - 1100nm

Wavelength Accuracy

±0.3nm

Wavelength Repeatability

≤0.2nm

Photometric Accuracy

0.2%T(0-100%T), ±0.002A(0-0.5A), ±0.004A(0.5-1A)

Photometric Repeatability

≤0.15%T(0-100%T), 0.001A(0-0.5A), 0.002A(0.5-1A)

Stray Light

≤0.03%T@220nm,360nm

Stability

±0.002A/h@500nm

Photometric Range

0-200%T, -0.3-3.0A, 0-9999C(0-9999F)

Baseline Flatness

±0.002A(200-1000nm)

Noise

0.0003A@500nm

Working Mode

T,A,C,E

Wavelength Setting

Automatic

Scanning Speed

High, Medium and Low

Detector

Solid Silicon Photodiode

Light Source

Tungsten Halogen/Deuterium Lamp

Data Output

RS232, USB(A), USB(B)

Processor

Cortex_M3, 120Mhz

Power Requirements

AC 110-220V 50-60Hz

Shipping Dimensions & Weight

770*630*340mm, 27kg

940*740*510mm

52KG

770*630*340mm, 27kg

940*740*510mm

52KG




Go Back
Print
More
Recommended News
More
Recommendations
[Upward]